+49 (089) 18 93 668 0 - Info@ScienceServices.de   We are here for you! Call us or request a call back here.
Language

EVE™ Automatic Cell Counter

NE-EVE-MC

The EVE™ automatic cell counter uses state-of-the-art optics and image analysis to automate cell counting. EVE™ is a bench-top counter, designed to measure cell count and viability (live, dead and total cells) accurately and precisely, using the standard trypan blue technique.

EVE™ is the succeeding model of the established Countess® automated Cell Counter.

Product Details

Description

Dieses Produkt ist nur für den Laborgebrauch für Forschungs- und Entwicklungszwecke bestimmt. Es ist nicht als Medizinprodukt (93/42/EWG,
2007/47/EG) oder In-Vitro-Diagnostika (Richtlinie 98/79/EG) zugelassen.

Key features and benefits

Application for a broad range of cell sizes and types

EVE™ counts primary (tissue, blood) cell lines as well as stem cells in a size range of 5-60µm and a concentration range of 1x10^4 - 1x10^7cells/ml.

Validated Cell Lines

Differentiation of clumped cells

Provides accurate counting results. It features an advanced analysis algorithm for proper counting of clumped cells.

Correct Counting of Clumped Cells

Cell size gating

Select a range of cell sizes included in counting using the gating function.

Cell Size Gating Function

Data storage and analysis

Store 500 test results on the internal memory, transfer the data to a PC by USB drive. An additional PC based software provides an advanced recounting function. Detailed image analysis allows to check counting results manually in-depth (down to every single count).

 

User friendly

7 inch LCD touch screen, bench top size, NO maintenance needed.

Convenient Use

EVE User Interface

Trypan blue staining

Trypan blue is a diazo dye, a vital stain used to selectively color dead cells blue.
Live cells or tissue with intact cell membranes are not coloured. Dead cells are shown as a distinctive blue colour as the stain can traverse their membranes.

Trypan Blue Staining

Analysis Time <20 sec./Test
Cell Concentration Range 1x10^5 - 4x10^6 cells/ml (optimum)
Cell Size Range 5-60µm
Loading Sample Volume 10µl/Test
Measuring Range 1x10^4 - 1x10^7 cells/ml
Staining Method Trypan Blue Stain
Data Export USB Drive
Electrical Supply 240V, 50-60Hz
Dimensions 27(L) x20(W)x9(H)cm

More Information

More Information
Manufacturer
NanoEnTek
Cell Size Range 5 - 60µm
Dimensions 27(L) x 20(B) x 9(H)cm
Electrical Supply 240V, 50-60Hz
Packing Unit 1 Stück
Weight 0.000000